1) Do John Doty's global calibration, computing correction factors $m$ and $c$ in the equation:
$$ correct\_value = m \cdot set\_value + c$$Make sure that $m$ is $\approx 1$ to 3 decimal places, and that $c$ is in $[-10,10]$
Print the value
2) Run heater calibration tests using the heater calibration test rig. Make sure that set heater current and measured heater current from housekeeping are within $[-4,4]$ mA. Also plot errors for visual inspection.
3) Run Ed's RTD check using the RTD check test rig. This makes sure that the measured RTD values are within 5% of theoretical values derived by Ed. Also plot the errors for visual inspection. These are for the platinum RTDs only, other tests will have to be contrived when we hook the CCDs up.